Shanghai Xinzhuang Instrument Co., Ltd. is a leading company in the field of laboratory equipment, dedicated to technological innovation and continuous improvement. With a strong focus on R&D, production, and sales, we are committed to delivering high-quality products and excellent customer service. Our mission is to support the development of scientific research and industrial applications through advanced solutions.
**Common Troubleshooting for Small Spray Dryers**
Here are some common issues you may encounter with your small spray dryer, along with possible causes and solutions:
1. **Fan not working**
- Possible cause: Inverter damaged or fan motor failure.
- Solution: Contact our technical support team for assistance.
2. **Electric heater not functioning**
- Possible cause: Fan not started, solid-state relay damage, or heater failure.
- Solution: Start the fan first, check the relay, or contact us for replacement parts.
3. **Air compressor not operating**
- Possible cause: Intermediate relay failure, compressor not starting, or compressor damage.
- Solution: Check the relay, ensure the compressor is properly activated, or contact us for repair.
4. **Equipment not powered**
- Possible cause: Unstable power supply, air switch 1 off, or relay failure.
- Solution: Verify the external power source, turn on air switch 1, and check the relay.
5. **Touch screen not displaying**
- Possible cause: Start button damage or 24V switching power supply failure.
- Solution: Replace the start button or power supply, and contact us if needed.
6. **PLC not working**
- Possible cause: Air switch 2 is off or PLC module failure.
- Solution: Turn on air switch 2 or contact our support for further assistance.
7. **Inlet air temperature not displayed**
- Possible cause: Loose PT-100 probe, damaged PT-100 sensor, or temperature module issue.
- Solution: Contact Nanjing Pusen Instrument Equipment Co., Ltd. for professional help.
8. **Outlet air temperature not displayed**
- Possible cause: Similar to above—loose connection, damaged probe, or faulty module.
- Solution: Reach out to Nanjing Pusen for diagnosis and repair.
9. **Inlet air temperature not reaching set value**
- Possible cause: Excessive fan airflow.
- Solution: Adjust fan parameters to optimize performance.
10. **Low inlet air temperature at the bottom of the drying chamber**
- Possible cause: Low atomizing air pressure, compressed air leakage, or incorrect peristaltic pump settings.
- Solution: Increase air pressure (2–3 Bar), check for leaks, and adjust pump parameters accordingly.
11. **Nozzle not working**
- Possible cause: Air valve not open, low pressure, oversized needle parameter, or solenoid valve failure.
- Solution: Open the valve (φ4 white air tube), adjust pressure, modify needle parameters, or replace the solenoid valve.